Specifications
BiPAP® FOCUS™

Product Specification
Pressure: CPAP 4 to 20 cm H2O (4 to 20 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ±5 cm H2O (±5 hPa)
  EPAP 4 to 20 cm H2O (4 to 30 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ±5 cm H2O (±5 hPa)
  IPAP 4 to 30 cm H2O (4 to 20 hPa)
Resolution: 1 cm H2O (1 hPa)
Dynamic Accuracy: ±5 cm H2O (±5 hPa)

Rise Time : Settings 1 to 6
(where 1 = 0.1 sec and 6 = 0.6 sec)
Resolution: 1
Accuracy: ±10% of setting (at R=50, C=20)
Rate: 1 to 30 BPM
Resolution: 1 BPM
Accuracy: ±1BPM or ±10 % of setting
whichever is greater over a 4 minute period.
I-Time and Rate settings cannot allow I-Time
to exceed expiratory time.
I-Time: 0.5 to 3 sec
Resolution: 0.1 sec
Accuracy: ±0.1+10% of setting sec
I-Time and Rate settings cannot allow
I-Time to exceed expiratory time.
Ramp Time: 0 to 45 min
Resolution: 5 min
Accuracy: ±10% of setting
Ramp Start: 4 cm H2O to EPAP (4 hPa to EPAP)
Resolution: 1 cm H2O (1 hPa)
Patient Parameters
Patient
Circuit Pressure:
Range: 0 to 35 cm H2O (0 to 35 hPa)
(continuous bar graph display)
Rate: Range: 0 to 99 BPM
Estimated Vt: Range: 0 to 4000 mL
Estimated MV: Range: 0 to 99 L/min
Patient Leak: Range: 0 to 200 L/min
Apnea Rate: Range: 0 to 99 per hour
Alarms
High System Pressure: 5 cm H2O (5 hPa) over target IPAP pressure
(Automatic)
Low System Pressure5: 5 cm H2O (5 hPa) under target EPAP pressure
(Automatic)
Apnea: Range: 20, 40, 60 sec or Off
Apnea Rate: Range: 5, 10, 20 apneas per hour or Off
Nurse Call: Supports normally open or normally closed systems
Physical
Dimensions: 5.5" H x 11.4" W x 14" D
(140 mm x 290 mm x 355 mm)
Weight: 10 lb (4.5 kg)
Power
Input Rating: 100VAC / 230VAC and 50/60 Hz
Battery Backup: 3.8 amp-hour (NiMH)
Fuse: No replaceable fuse
Nurse Call: Rated current: 0.280 A
Rated voltage: 250 V